Square-law instrument scales
θ: pointer deflection · I: current magnitude · k: scale constant. The same relationships apply to voltage for a voltage-square-law instrument.
- θ = kI²
- k = θcal / Ical²
- I = √(θ/k)
- θ₂/θ₁ = (I₂/I₁)²
- 1Find k from a known calibration point.
- 2Square the current ratio for forward calculation; take the square root for inverse calculation.
If 90° corresponds to 3 A, k = 90/9 = 10°/A². At 2 A the pointer moves 40°. Check: √(40/10) = 2 A.
- Equal current increments do not create equal deflection increments.
- This pointer square law is different from the DP-to-flow square-root conversion.
Pass Your Red Seal’s lessons and practice are independent study material. Source: Yokogawa: DP flow square-root feature
Try one
A square-law ammeter shows a deflection of 90° when carrying 3 A. What deflection will it show when the current is 1 A?